"CrystalGuard" - Monitoring System for Isolated Limb Perfusion
1. Description
CrystalGuard
CrystalGuard with Crystal Probe -automatic-
- the safe solution for high-risk chemotherapy - the best way for continuous isolation control; radionuclide based monitoring
- permanent validation of isolation during treatment
- small detector size prevents interferences
- signal acquisition and processing is provided by our control unit CXS-SG04
- battery or mains operated
- visual and acoustical alarm
- data visualisation and storage by a PC
2. Specifications
- rugged solid state sensors with high efficiency
- Size: D= 70mm; height 30mm, integrated electronics
- Housing: water proof
- Connection: cable, 3m
- Energy range: 80 keV up to 1 MeV, optimized for 140 keV (Tc-99m-gamma-radiation)
- Sensitivity: 200.000 counts/s/MBq @140 keV
- Max. count rate: 105 counts/s
- Scintillation crystal: CsI:Tl, D 25.4x10 mm³, and photodiode-read-out
- Energy resolution: about 25%@ 140 keV
- Side shielding: 4 mm Pb from all sides; radiation entrance window on the bottom of the detector (shielding > 99.97% @ 140 keV)