"CrystalGuard" - Monitoring System for Isolated Limb Perfusion

1. Description

if you can read this, then something went wrong.

CrystalGuard


if you can read this, then something went wrong.

CrystalGuard with Crystal Probe -automatic-

  • the safe solution for high-risk chemotherapy - the best way for continuous isolation control; radionuclide based monitoring
  • permanent validation of isolation during treatment
  • small detector size prevents interferences
  • signal acquisition and processing is provided by our control unit CXS-SG04
  • battery or mains operated
  • visual and acoustical alarm
  • data visualisation and storage by a PC

2. Specifications

  • rugged solid state sensors with high efficiency
  • Size: D= 70mm; height 30mm, integrated electronics
  • Housing: water proof
  • Connection: cable, 3m
  • Energy range: 80 keV up to 1 MeV, optimized for 140 keV (Tc-99m-gamma-radiation)
  • Sensitivity: 200.000 counts/s/MBq @140 keV
  • Max. count rate: 105 counts/s
  • Scintillation crystal: CsI:Tl, D 25.4x10 mm³, and photodiode-read-out
  • Energy resolution: about 25%@ 140 keV
  • Side shielding: 4 mm Pb from all sides; radiation entrance window on the bottom of the detector (shielding > 99.97% @ 140 keV)